| REFERENCE HYGROMETER |
| Type |
Optidew cooled mirror dew-point transmitter |
| HUMIDITY |
| Generator Range |
10-90% RH |
Accuracy Control Element |
≤±1% RH (10–70% RH) ≤±1.5% RH (70–90% RH) |
| Stability |
±0.2% RH (20–80% RH) |
| TEMPERATURE |
| Generator Range |
+10°C to +50°C (lowest T set point = 10ºC below ambient) |
| Accuracy |
±0.1°C |
| Stability |
±0.1°C |
| CHAMBER |
| Ramp Rate |
+20 to +40°C: 1.5°C/minute +40 to +20°C: 0.7°C/minute |
| Control Element |
Removable relative humidity sensor |
| GENERAL |
| Probe Ports |
Up to 5 – sensor body diameters 5 – 25mm accomodated by port adapters |
| Chamber Volume |
2000cm3 |
| Chamber Dimensions |
105 x 105 x 160mm (w x h x d) |
| Instrument Dimensions |
520 x 290 x 420mm (w x h x d) |
| Set Point Resolution |
0.1 for humidity and temperature |
| Displays |
High definition 2 lines alpha numeric |
| Measurement Units |
ºC, ºCdp, % RH g/kg, g/m3, (t-td), aw |
| Outputs |
Analogue: Two channels 4-20 mA or 0-20 mA Digital: RS232 (RS485 optional) Alarm: Volt free contact, 2A @ 30 V DC |
| Supply |
85–264 V AC, 47–63 Hz, 150 VA |
| Weight |
20kg |
|